How to make the calibration error of the electronic balance can be reduced to the minimum

<

In the verification (test), we found that the first measurement test of the balance has a large error. The reason is that a considerable part of the instrument has not been calibrated in a long time interval, and it is considered that the balance can be directly weighed by displaying the zero position. . (It should be pointed out that the electronic balance starts to display zero point. It cannot be said that the accuracy of the balance weighing data meets the test standard, and only the balance zero stability is qualified. Because it is necessary to consider the other technology when measuring whether a balance is qualified or not. Compliance of indicators). Due to long storage times, positional movements, environmental changes or for accurate measurements, the balance should generally be calibrated prior to use. The calibration method is divided into internal calibration and external calibration. The German-produced Saudilis, the Swiss-made METTLER, and the AL-series analytical balances produced by the factory have an average calibration device. If you do not read the instructions carefully before use, it is easy to ignore the "calibration" operation, resulting in a large weighing error.

How to externally calibrate the balance. Method: Press the CAL button when the CAL- appears on the display, the display will appear CAL-100, where “100” is the flashing code, indicating that the calibration weight requires a standard weight of 100g. At this point, put the "100g" calibration weight on the weighing plate, and the display will appear "----" waiting state.

After a long time, the display will appear 100.000g, and the calibration weight will be taken. The display should appear 0.000g. If it is not zero, then clear it again and repeat the above calibration operation. (Note: In order to get accurate calibration results, it is best to repeat the above calibration procedure twice.)

Methods as below:
Set the balance to zero, then keep pressing the “CAL” button until CALint appears. The following conditions will be displayed during calibration:
Balance zero
Internal calibration weight loaded
Balance rechecking zero
Balance report calibration process
The balance report is calibrated
The balance automatically returns to the weighing state
Some people think that the heavier the object weighed within the range of the electronic balance, the greater the damage to the balance. This understanding is not entirely correct. The maximum safe load of a general weighing instrument is the maximum static load that it can withstand without causing a permanent change in its metering performance. Electromagnetic force automatic compensation circuit principle, when the weighing pan is loaded (be careful not to exceed the weighing range), the electromagnetic force will push the weighing pan back to the original equilibrium position, so that the electromagnetic force is balanced with the gravity of the object, as long as it is allowed The effect of the weighing scale on the balance is small and does not affect the accuracy of the electronic balance due to long-term weighing.